Specimen Requirements

Collection Requirements Before collection, platelet study must be scheduled with Coagulation laboratory at 214-633-4959. Platelet count should be greater than 100,000. Specimens should be collected in 2-3 mL 3.2% sodium citrate (light blue top) tubes. Do not under or over-fill collection tubes. Gently invert the tube end-over-end 8-10 times. Do not place in tube system.
Shipping / Handling Instructions Specimen must remain at room temperature. Specimen must be received within 2 hours of collection. Specimenmust be accompanied by a “Platelet Aggregation Study History Form”. The intrepretation may be delayed until form has been submitted.

Container Type

Laboratory Container Default Temperature Minimum Volume
CUH LAB 5 BLUES 1 PURPLE Yes Room Temperature

Performing Section

CUH LAB: CUH COAG

Components

Component Name Component ID LOINC Code
10.0 UM ADP HYPERACTIVE PLT 1230000384
5.0 UM ADP HYPERACTIVE PLT 1230000390
2.5 UM ADP HYPERACTIVE PLT 1230000388
1.25 UM ADP HYPERACTIVE PLT 1230000386
2.0 MCG/ML COLLAGEN HYPERACTIVE PLT 1230000389
1.0 MCG/ML COLLAGEN HYPERACTIVE PLT 1230000387
0.5 MCG/ML COLLAGEN HYPERACTIVE PLT 1230000382
0.25 MCG/ML COLLAGEN HYPERACTIVE PLT 1230000379
0.500 MM ARACHIDONIC ACID HYPERACTIVE PLT 1230000380
0.250 MM ARACHIDONIC ACID HYPERACTIVE PLT 1230000378
0.125 MM ARACHIDONIC ACID HYPERACTIVE PLT 1230000385
0.0625 MM ARACHIDONIC ACID HYPERACTIVE PLT 1230000383
INTERP HYPERACTIVE PLT 1230002322